Edited by: Andre Lin Ouedraogo, Bill and Melinda Gates Foundation, United States
Reviewed by: Kingsley Badu, Kwame Nkrumah University of Science and Technology, Ghana; Jana Held, University of Tübingen, Germany
*Correspondence: Roxanne R. Rees-Channer,
This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.
Light microscopy remains a standard method for detection of malaria parasites in clinical cases but training to expert level requires considerable time. Moreover, excessive workflow causes fatigue and can impact performance. An automated microscopy tool could aid in clinics with limited access to highly skilled microscopists, where case numbers are excessive, or in multi-site studies where consistency is essential. The EasyScan GO is an automated scanning microscope combined with machine learning software designed to detect malaria parasites in field-prepared Giemsa-stained blood films. This study evaluates the ability of the EasyScan GO to detect, quantify and identify the species of parasite present in blood films compared with expert light microscopy.
Travelers returning to the UK and testing positive for malaria were screened for eligibility and enrolled. Blood samples from enrolled participants were used to make Giemsa-stained smears assessed by expert light microscopy and the EasyScan GO to determine parasite density and species. Blood samples were also assessed by PCR to confirm parasite density and species present and resolve discrepancy between manual microscopy and the EasyScan GO.
When compared to light microscopy, the EasyScan GO exhibited a sensitivity of 88% (95% CI: 80-93%) and a specificity of 89% (95% CI: 87-91%). Of the 99 samples labelled positive by both, manual microscopy identified 87 as
This study shows that the EasyScan GO can be proficient in detecting malaria parasites in Giemsa-stained blood films relative to expert light microscopy and accurately distinguish between
香京julia种子在线播放
Current WHO estimations suggest that malaria is responsible for over 200 million infections annually, of which approximately half a million cases lead to death (
Automated malaria diagnosis systems have several potential benefits: they do not fatigue; they give reproducible results; they can examine greater quantities of blood to give more stable results; they can increase the productivity of overworked technicians and pathologists; and they can be widely deployed, addressing the expert-training bottleneck. For example, automated systems (if sufficiently accurate) would be well-suited to drug resistance monitoring (
These benefits of automated systems are offset by drawbacks (
In addition to software algorithms, an automated system requires reliable hardware to scan blood slides and capture images for analysis. Because of the blood volume required (e.g. to contain 500 white blood cells at a minimum (
In this study, we tested a fully automated malaria diagnosis system that combines ML software developed by Global Health Laboratories (GHL) with the EasyScan GO, an automated scanning microscope developed by
Returned travelers who have recently visited malaria-endemic countries and are unwell are routinely assessed for possible malaria infection in clinics at The Hospital for Tropical Diseases and Homerton University Hospital, London, UK. Diagnostic evaluation primarily involves examination of patient blood samples by manual light microscopy (using stained peripheral blood smears) to confirm the presence of malaria parasites, determine the species and provide an estimation of parasitemia where
Thick and thin smears were prepared on clean glass slides using surplus patient EDTA blood samples which had been obtained by venepuncture (
For each blood sample, there were two independent reads performed. The first result was obtained from the routine diagnostic laboratory after examination by two microscopists within that laboratory and was used to determine if samples were malaria parasite positive or negative for recruitment purposes. The second reading was study-specific, also provided by an expert microscopist, who examined slides produced from the same blood samples after they had been anonymized to confirm positivity and perform accurate quantitation. If there was a discrepancy between the results provided by the study microscopist and the initial diagnostic laboratory microscopy result, a second expert study microscopist was then engaged to perform a third read.
The EasyScan GO is a fully automated end-to-end malaria diagnostic system which includes both hardware and software. An automated bright-field microscopy platform scans Giemsa-stained thick and thin blood films, and malaria detection algorithms process the image sets to give parasite detection, species ID, and parasite quantitation for the patient. Given a blood film, the device automatically scans and processes the slide, and outputs a report that includes estimated diagnosis, species ID, quantitation, WBC count, and a mosaic of thumbnail images of top suspected malaria parasites (
A typical patient thick film report for a
Thick films are used to
Briefly, a new sample is assessed as follows: In a thick film the algorithm analyses 100 image stacks (each 113 µm x 85 µm x 5 depths), containing an average of 1527 WBCs (std dev 862; 90% of samples have over 720 WBCs). Candidate parasite objects are detected, then culled, by rapid morphological methods, which are tuned for high sensitivity yet still eliminate most of the easier distractor objects. The remaining candidate objects pass through two convolutional neural net (CNN) classifiers to receive labels as ring, late-stage, or distractor. WBCs are detected and counted by a separate module. Diagnosis is based on whether the suspected parasite count per WBC exceeds a noise threshold that has been pre-tuned to aim for 90% patient specificity. While the high blood volume examined theoretically allows for a lower limit of detection, in practice the object false positive rate is a limiting factor (
Training slides included over 500 imaged blood slides from 12 countries, encompassing a wide range of different staining presentations and containing a variety of artefacts. The collection included large numbers of
The WHO international standard for
For real-time PCR, in the first instance, each sample was amplified in a multiplex reaction targeting the conserved region of the
Nested PCR (nPCR) is widely regarded as the gold standard nucleic acid amplification (NAA) method for detection of malaria parasites in very low density samples (
The required sample size of 104 malaria positive slides and 1125 malaria negative slides was calculated as a non-inferiority study to be able to jointly test that the sensitivity of EasyScan GO was not decreased by more than 80% compared to expert microscopy, and that the false positive fraction (1-specificity) was not increased by more than 50% compared to expert microscopy, with 5% significance and 80% power. The malaria positive slides were expected to be derived from symptomatic clinical cases, plus follow-up slides from these same patients following treatment (i.e. with lower parasitemia). The sensitivity and specificity for expert microscopy with this slide composition was assumed to be 80% and 90%, respectively.
A total of 1202 patient samples were collected and the same sets of Giemsa-stained thick and thin slides evaluated concurrently by manual light microscopy and the EasyScan GO. By light microscopy, 113 of the samples were confirmed as malaria parasite positive and 1089 were negative. When compared to light microscopy, the EasyScan GO exhibited a sensitivity of 88% (95% CI: 80-93%) and a specificity of 89% (95% CI: 87-91%). The EasyScan GO correctly identified 99 of the 113 light microscopy positives but also incorrectly reported a positive result for 122 samples that were identified as parasite negative (1089) by light microscopy (
Diagnostic accuracies of
(A) LM vs PCR | (B) EasyScan GO vs PCR | (C) EasyScan GO vs LM | |
---|---|---|---|
Sensitivity, % (95% CI), N/Ntot | 72% (64-79), 111/154 | 69% (61-76), 106/154 | 88% (80-93), 99/113 |
Specificity, % (95% CI), N/Ntot | 100% (99-100), 1046/1048 | 89% (87-91), 933/1048 | 89% (87-91), 967/1089 |
Likelihood ratio (+), N (95% CI) | 377.69 (94-1513) | 6.27 (5.12-7.68) | 7.82 (6.53-9.37) |
Likelihood ratio (-), N (95% CI) | 0.28 (0.22-0.36) | 0.35 (0.28-0.44) | 0.14 (0.09-0.23) |
Of the 99 samples labelled positive by both light microscopy and the EasyScan GO, manual microscopy identified 87 as
To achieve level 1 (expert) competency in malaria microscopy, the WHO requires that 50% of samples containing malaria parasites with densities between 200 and 2000 p/uL be quantified within +/-25% of the “true count” (
Comparison of parasite densities estimated by EasyScan GO vs manual light microscopy. Dotted green lines correspond to +/- 25% error. PCR-positive samples are black circles (Pf) or blue triangles (non-Pf). PCR-negative samples are red circles.
Blood film examination for malaria parasites is far from extinct, despite the wide uptake of malaria rapid diagnostic tests (RDTs). Indeed, the 2022 edition of the British Society for Haematology guidelines for the laboratory diagnosis of malaria states “Rapid diagnostic tests (RDTs) for malarial antigen cannot replace microscopy but can be useful as a supplementary test when malaria diagnosis is performed by relatively inexperienced staff. They should not be used instead of a film at any time including out of hours” (
In this study, diagnostic accuracy of the EasyScan GO was similar to that of expert Light Microscopy. The principal difference was the EasyScan GO’s lower specificity (89% vs 100%, PCR as reference). From a clinical perspective, false positives could mean patients receiving drug treatment that is not required, therefore the device would not be used as a standalone diagnostic tool without additional input from other laboratory staff needing to screen further for potential false positives. However, the mosaic of thumbnails of suspected parasites (e.g. as shown in
The Easy Scan GO accurately distinguished
Evolving drug resistance (
An important detail to note is that light microscopy parasite quantitations in this study were based on accurate total WBC counts which varied widely by individual (mean 5430, std dev 2070), while EasyScan GO assumed a fixed value of 8000 WBCs/µL. When correcting parasitemia estimates provided by EasyScan GO using accurate individual WBC counts, the percentage of samples having quantitation error within the 25% margin increases to 50%. This indicates that quantitation error was affected by how WBC counts per µL were defined for the samples examined.
The EasyScan GO system, and very similar algorithms deployed on different hardware, have been evaluated in four other field trials (
1. The same EasyScan GO system was applied to a WHO 55 reference set and had somewhat stronger performance vs PCR (87% sensitivity, 100% specificity) (
2. The same EasyScan GO system, minus the thin film algorithm and applied to thick films only, was also applied to field slides (170 Light Microscopy (LM)-positive, 623 LM-negative) from a variety of sites in Thailand and Indonesia (
3. A system with a very similar thick film algorithm (no thin film algorithm) and a different scanning microscope (the Autoscope, not the EasyScan GO) was tested on field samples (thick films only) in Peru (
4. The same system as used in the present study, minus the thin film module and applied to thick films only, was tested on field samples at 11 sites in 11 countries (
5. Two internal algorithm assessments reported similar though somewhat more optimistic results (
The perspective afforded by five separate field trials is, to our knowledge, unique for an automated malaria diagnosis system. This perspective is highly valuable because a system’s performance can vary in different settings. Since machine learning-based systems can be brittle in the face of new data sources, the high variability of slide preparation at different clinics is a serious challenge for automated malaria diagnosis systems. In this context, therefore, multiple data points on system performance are especially important to understanding a system’s suitability for deployment. Collectively, these trials demonstrate strong performance of a fully automated system for assessing a diversity of Giemsa-stained blood films.
Manual malaria microscopy requires significant expertise and even expert microscopists become fatigued in the face of a heavy workload, with the potential for error. An automated system such as the EasyScan GO would have the capacity to reduce workload for individual microscopists whilst retaining the option for a technician or pathologist to quickly check the device’s findings using the mosaic of thumbnails of suspected parasites it produces. This reflects the general fact that, currently, automated Machine Learning systems do not match the capabilities of expert microscopists at malaria diagnosis.
In this study, the EasyScan GO fell short in performance compared to that of expert manual light microscopy in terms of sensitivity and specificity (88% and 89% respectively). The EasyScan GO wrongly identified 122 samples as positive that were read as parasite negative by light microscopy. However, as mentioned above, this limitation could be partially mitigated by the output of thumbnail images of these wrongly identified parasites that can be rechecked by a microscopist on site or from a remote location. From all malaria positives identified by both light microscopy and the EasyScan GO, the latter accurately identified all but one
The dependence of quantitation accuracy on the ground truth method for counting WBCs/µL suggests possible future paths for automated microscopy: for example, scanning a known volume of blood might improve quantitation accuracy e.g. the Earle and Perez method, that does not require a microscopist to manually count WBCs in order to estimate parasitemia accurately (
As machine learning advances further and has the opportunity to learn from exposure to more positive malaria sample images as well as a wide range of background and staining artefacts, fully automated systems such as EasyScan GO will have a future in malaria diagnosis in a variety of settings in both endemic and non-endemic areas.
The raw data supporting the conclusions of this article will be made available by the authors, without undue reservation.
The studies involving human participants were reviewed and approved by London-Central Research Ethics Committee. Written informed consent for participation was not required for this study in accordance with the national legislation and the institutional requirements.
CB and PC assisted with study design. KW and PC were responsible for identifying and recruiting study participants. RR-C, LG, PL, KB and SS were responsible for SP development and laboratory work. MH, CD, LH, CM, and CT developed the software algorithm. Study data was analysed by MG, MH, SB, CD and RR-C. CB, CD, RR-C and PC were responsible for drafting the manuscript. All authors contributed to the manuscript and approved the final version.
All funding for the study was provided by The Global Good Fund I, LLC.
The authors would like to thank all collaborating staff within Research Offices at the London School of Hygiene and Tropical Medicine, Homerton University Hospital NHS Foundation Trust, and University College London/University College London Hospitals NHS Foundation Trust (the Joint Research Office). Specifically, we acknowledge the invaluable assistance provided by senior research nurses Monica James (Homerton) and Michelle Berkeley (UCL/UCLH). In addition, we would like to thank laboratory staff within the department of Haematology at Homerton University Hospital NHS Foundation Trust for the preparation and provision of blood smears for their patients enrolled in the study.
Author CT was employed by the company Creative Creek, LLC.
The remaining authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.
All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.
The Supplementary Material for this article can be found online at: